Return To Table of Contents
24
2141-209-002 REV H
www.stryker.com
Maintenance Menu Guide
Figure 2: Configuration Screen
Figure 1: Footboard Control Panel
Brake
HOB
Drive
Main Menu
The Maintenance Menu is accessed through the Touch Screen and contains additional features of the product. This
menu provides an interface to the user and/or service personnel in order to provide the ability to control and access
maintenance features.
ACCESSING THE CONFIGURATION SCREEN
WARNING
Please ensure patient is not in the bed prior to starting bed calibration. In calibration mode, the software does not control
the interferences between the mechanical parts of the bed. Mechanical damage could occur without supervision. Only
qualified personnel should perform the calibration.
Note
Verify that the bed is on a level surface which does not have any slopes or inclines prior to entering into the calibration
mode.
1. Unseat and reseat the footboard and wait until the main control screen is displayed.
2. Push and hold the Main Menu button located in the upper right corner of the footboard control panel
(see Figure 1 below). Continue pressing on the Main Menu button while executing steps 3-5.
3. Push and hold the
Brake
button for 5 seconds and then release.
4. Push the
HOB 30
o
+
button once and release.
5. Push the
Drive
button once and release.
6. Release the Main Menu button and you will be taken into the Configuration Screen shown in Figure 2.