Maintenance Menu Guide
RUNNING A DIAGNOSTIC TEST FOR ISOLIBRIUM
WARNING
Do not perform diagnostic test with a patient or other weight on the support surface.
CAUTION
Do not touch the support surface while performing diagnostics test. Failure to do so could lead to inaccurate diagnostic
results.
Notes
• The results of the test will not be displayed until all tests have been completed.
• No results will be displayed if the tests are canceled.
• Other diagnostic functions are disabled while the diagnostic test is running.
The diagnostic test runs seven tests at one time on the support surface and take approximately 45 minutes to complete.
They include the following:
• Seat Left
• Seat Right
• Back Left
• Back Right
• Turn Left
• Turn Right
• Low Air Loss
To run diagnostics, tap
start
on the lower portion of the Diagnostics screen (Figure 46).
Figure 46: Diagnostic Main Screen - Sample