Return To Table of Contents
42
2141-209-002 REV H
www.stryker.com
Note:
For
Isolibrium
error codes, see the
Isolibrium
operations/maintenance manual.
To enter the
Active Errors
screen, tap the
Active Errors
tab
on the Service Diagnostic screen (Figure 42).
A list of the active errors are displayed (Figure 42). A maximum of five active errors are displayed simultaneously, and
are ordered by priority.
REVIEWING AND CLEARING ISOLIBRIUM ERROR HISTORY
To enter the
Error History
screen, tap the
Error History
tab on the Diagnostic Screen (Figure 43).
The error list is ordered from the oldest entry to the newest, with a maximum of 10 errors.
To clear the
Error History
screen, tap
Clear Error History
on the Diagnostic screen (Figure 43).
Notes
• Always tap
Clear Error History
after troubleshooting is complete
• If you clear the error history while there is an active error, the active error does not show in the
Error History
tab
until the active error is fixed.