Preventive Maintenance
All grounds secure to the frame
Ground impedance not more than 100 mΩ (milliohms)
Current leakage not more than 300 μA (microamps)
Ground chains are clean, intact, and have at least two links touching the floor
Enclosure is free from wear, tear, stresses and mechanical damage
No rust or corrosion of parts
Labels for legibility, proper adherence, and integrity
iBed Wireless Module and IR Module are intact and footboard icons are displaying (iBed Wireless option)
Serial Number:
Completed by: _______________________________________
Date: _________________