INSTALLING THE XPRT THERAPY SUPPORT SURFACE (OPTIONAL)
To install the XPRT support surface option onto InTouch, see the installation instructions in the XPRT support surface
operations manual.
INSTALLING THE POSITIONPRO SUPPORT SURFACE (OPTIONAL)
To install the PositionPRO support surface option onto InTouch, see the installation instructions in the PositionPRO
support surface operations manual.
INSTALLING THE ISOLIBRIUM SUPPORT SURFACE (OPTIONAL)
To install the
Isolibrium
support surface option onto InTouch, see the installation instructions in the
Isolibrium
support
surface operations manual.
INSTALLING
I
BED WIRELESS (120V NORTH AMERICAN ONLY) (OPTIONAL)
WARNING
• The optional iBed® Wireless function is only intended to provide remote information of product status and parameter
conditions. It is not intended to replace patient monitoring protocol.
• The line of sight between an iBed Locator and the head end of bed must be free of obstruction at all times. Any
interference could interrupt communication between the iBed Locator and the IR module.
• You must use an iBed® Wireless compatible footboard with an iBed® Wireless compatible product. You will lose
iBed® Wireless functionality if you use an older version of the footboard.
• You must verify all iBed® Wireless functionality after installation. Verify that the iBed locator communicates the
product’s positions, and that iBed® Wireless communicates with the wireless access point. Failure to do may result
in the loss of information or the transmission of incorrect information.
• You must install the iBed Locators more than 71” apart from one another in the same room (such as in a semi-
private room with more than one product). Failure to do may result in the product transmitting information to the
incorrect iBed Locator.
• Do not use iBed® Wireless to replace the existing nurse call system. iBed® Wireless products are only intended to
transmit product information. They are not intended to transmit nurse call information.
You must install the
i
Bed Locator on the wall at the head end of the bed. The
i
Bed Locator communicates with the IR
Module that is installed onto the product.
To mount the
i
Bed Locator, see the installation instructions that were included with your
i
Bed Locator Installation kit.
After you install the
i
Bed Locator, you must configure the wireless connection settings for the
i
Bed Server application.
To configure the
i
Bed Server application, see the installation and configuration instructions in the
i
Bed Server
Installation/Configuration manual.
WARNING
• Always correctly associate or map the
i
Bed Locator to the room or location to provide accurate location information.
Failure to properly map the
i
Bed Locator to the room or location yields incorrect remote information.
• If you move an
i
Bed Locator after it has been installed and mapped, you must remap to the new room or location
in which it is moved to. You must also remap the
i
Bed Locator if you change the room or location information after
initial installation.
If you have any problems during the
i
Bed Wireless installation process, contact Stryker Technical Support at
1–800–327−0770.