Return To Table of Contents
www.stryker.com
2141-209-002 REV H
91
Service Information
Tools Required:
• Phillips Screwdriver #2
• Needle-Nose Pliers
• 3/8” Drive Ratchet
• 1/2” Socket
• 1/2” Combination Wrench
Procedure: (procedure the same for left and right sides)
1. Plug bed into wall outlet.
2. Raise bed and fowler to full up position, gatch down and foot up.
3. Raise head end siderails.
4. Using a Phillips screwdriver, remove the two screws securing the wire cover to the head frame (Figure 1).
5. Using a Phillips screwdriver, remove the six screws from the head frame cover, lift up on the cover and lay it on
the litter (Figure 2)
6. Unscrew the load cell connection and feed the connection back towards the cell.
7. Using Needle-nose pliers, remove the rue clip and the flat washer.
8. Using one hand, grab the upper frame cross bar to remove tension on the litter then push the clevis pin out.
9. Using a ratchet with a 1/2” socket and a 1/2 combination wrench, remove the two bolts securing the load cell then
remove the load cell.
10. Reverse procedures to install the new load cell.
CAUTION
Ensure cables do not get pinched when securing the head end frame cover together.
11. Recalibrate the bed (refer to Bed Calibration procedures on
page 25
).
12. Test all bed functionality prior to putting bed back into service.
LOAD CELL (HEAD END) REMOVAL AND REPLACEMENT - (LITTER)
Head Frame Cover
Head Frame
Screw
Screw
Screw
Screw
Screw
Screw
Upper Frame
Screw
Clevis Pin
Flat Washer
Wire Cover
Rue Clip
Figure 1
Figure 2