Maintenance Menu Guide
INTOUCH CONFIGURATION SCREEN (CONTINUED)
A. Bed Calibration (Continued)
13. When step 2 of the calibration procedure has completed, step 3 of the calibration procedure will begin and
Figure 13 as shown below will appear on the screen.
14. Place the bed at +12 degrees Trendelenburg by pressing
the
Trendelenburg (Foot Lift Up)
button as shown in
Figure 14. 12 d/- 0.1 with the inclinometer
you previously placed on the litter frame in step 11.
Figure 14: Head Siderail Control Panel
Figure 13: Bed Calibration - Step 3
QDF27-2237
Trendelenburg