Return To Table of Contents
www.stryker.com
2141-209-002 REV H
37
Maintenance Menu Guide
INTOUCH CONFIGURATION SCREEN (CONTINUED)
C. Touch Screen Calibration
Ensure patient is not present on the bed before performing the Touch Screen Calibration.
1. From the Configuration Screen, press the Touch Screen Calibration button. Figure 28 will appear.
NOTE: If the touch screen will not respond, you will need to push the HOB 30
0
+, Brake and Drive buttons
on the footboard at the same time (this will enable you to get directly into the touch screen calibration).
CAUTION: When pushing the buttons above the fowler, brakes, or drive, actuators may activate.
2. Carefully press and briefly hold a stylus or your finger on the center of the cross hair shown on the screen.
3. Repeat as the target moves around the screen. There will be five different locations to press: center, lower
right, upper right, upper left, and lower left.
4. When the last cross hair has been touched, Figure 29 displays ‘New calibration settings have been measured’.
5. Tap the screen to register saved data.
6. If you do not want to save the new data, wait for 30 seconds to cancel saved data and keep the current setting.
NOTE: The TOUCH SCREEN CALIBRATION SCREEN is a validation of the touch screen’s calibration. When the
user presses anywhere on the screen, the coordinates x and y are displayed on the lower part of the screen.
Figure 28: Touch Screen Calibration: Start
Figure 29: Touch Screen Calibration: Completed