Return To Table of Contents
36
2141-209-002 REV H
www.stryker.com
Maintenance Menu Guide
INTOUCH CONFIGURATION SCREEN
(CONTINUED)
B. Full Diagnostic
1. From the Configuration Screen, press the
Full Diagnostic button. Figure 26 will appear.
2. The following menu items may be selected
by pressing their button.
a.
BOARDS
(Figure 27a)
Provides information on the switch
boards and the touch screen’s
software version.
b.
ERROR CODES
(Figure 27b)
Provides information on errors which
the CPU board has identified.
c.
INPUT STATES
(Figure 27c)
Provides information on the status of
all switches and jumpers on the bed.
d.
MOTORS
(Figure 27d)
Provides information on what a motor is doing when a function button is pushed.
Note:
This requires assistance to press the buttons on the head siderails or at the head end control.
e.
BUTTONS PRESSED
(Figure 27e)
Provides information on when the CPU has detected a button being pressed.
Note:
This requires assistance to press the buttons on the head siderails or at the head end control.
f.
SIGNAL VALUES
(Figure 27f)
Provides information on CPU voltages, load cell values, and angle sensor values.
Figure 27a: Board
Boards Connected
Close
Nurse A.0 A.1
Nurse (SRH) A.1
Nurse (SRL) A.1
Nurse (SRR) A.1
Power board B.37
Room Interface A.1
TouchScreen D.3
Motors
Close
Gatch - Reverse
Hilo Foot - Forward
Foot - Forward
Buttons Pressed
Close
Nurse SRR - Trend
Footboard - Hob
Nurse Head - Leg up
Figure 27c: Input States
Figure 27b: Error Codes
Figure 27d: Motor
Figure 27e: Pressed Buttons
Figure 27f: Signal Values
Figure 26: Full Diagnostic Screen