TREND ANGLE SENSOR REMOVAL AND REPLACEMENT - (LITTER)
Tools Required:
• Phillips Screwdriver #2
• ESD System
Procedure:
1. Plug the bed into a wall outlet.
2. Raise the bed to the full up position and raise the patient right siderail.
3. Working from the patient’s right side, use a Phillips screwdriver and remove the two screws securing the trend
angle sensor to the litter frame (refer to Figure above),
4. Using an ESD system, properly ground yourself.
5. Unclip the three clips holding the board cover on.
6. Unplug the cable from the board.
7. Reverse the steps to install new trend angle sensor.
Note: Do not over-tighten the two screws.
8. Recalibrate the bed (refer to Bed Calibration procedures on
page 25
).
9. Test all bed functionality prior to putting the bed back into service.
Head
Foot
Litter Angle Sensor