9-2
HDW-750/750CE V1
Setup menu (PAINT menu) :
.
01
*
SW STATUS
GAMMA
→
ON
BLACK GAMMA
→
OFF
MATRIX
→
OFF
KNEE
→
ON
WHITE CLIP
→
ON
DETAIL
→
ON
APERTURE
→
ON
TEST SAW
→
OFF
.
03
*
BLACK
OUTPUT SELECT
→
Y
.
06
*
KNEE
KNEE SATURATION
→
ON
KNEE SAT LEVEL
→
0
.
10
*
SKIN DETAIL
SKIN DETAIL ALL
→
OFF
Setup menu (MAINTENANCE menu) :
.
14
*
FUNCTION 3
WHT FILTER INH
→
OFF
COLOR BAR SEL
→
100%
.
15
*
GENLOCK
GENLOCK
→
ON
RETURN VIDEO
→
OFF
9-1-4. Gray-Scale Chart and Its Maintenance
Use of a reflective-type gray-scale chart (with reflectance
of 89.9%) is recommended for the PA Gain adjustment. If
a reflective-type gray-scale chart is not available, use a
transparent-type gray-scale chart with a well-maintained
pattern box.
Calibrate illuminance (or luminance) on the surface of the
gray-scale chart to the specified value as follows. Also,
calibrate the color temperature to exactly 3200K by
adjusting the illumination.
Reflective-Type Gray-Scale Chart (4 : 3)
Availability of reflective-type gray-scale chart
The reflective-type gray-scale chart (4 : 3) is commercially
available.
Name :
Reflective-type gray-scale chart in
wooden case (GS-3 of Murakami Color
Research Lab. Inc. or equivalent)
Available from : Murakami Color Research Lab. Inc.
Address : 3-11-3 Kachidoki, Chuo-ku,
Tokyo 104-0054
TEL :
(Japan) 03-3532-3011
FAX :
(Japan) 03-3532-2056
Notes on handling
.
Do not touch the chart surface with bare hands.
.
Do not rub or stain the chart surface.
.
Do not expose or use the chart under direct sunlight for a
long time.
.
Do not leave or store the chart in a place that is very
humid or contains toxic gas.
.
Do not place anything on top of the chart surface.
.
When a chart is stored for a long time without using it,
open the wooden case and leave the chart to dry for 30
minutes or 1 hour once or twice every month.
Guideline of replacing the chart when used as the
reference chart
The reflective characteristics of the reflective-type gray-
scale chart can easily deteriorate over time. As a guideline,
the chart should be replaced every 2 years, but the replace-
ment interval depends on the storage conditions of the
chart.
9-1. General Information on Electrical Alignment