28
Section 2
Pump Programming - TPN Delivery Mode
TPN: Starting Daily Infusion
When a new administration set is attached to the pump at the beginning
of infusion, the reservoir volume should be reset. This will also reset the
infusion profi le. To do this, follow the instructions on page 36 for attaching
an administration set and answering yes to “Reset reservoir volume to ... ?”
Then when you start the pump, delivery will begin at the start of the infusion
period (illustrated below).
NOTE: Whenever the reservoir volume is reset, the infusion profi le is also reset so
the infusion will start at the beginning.
Infusion Peroid
Infusion Volume
Taper-Up
Period
Taper-Down
Period
KVO
Infusion Profile
Plateau Rate
(must be at least 10 min.)
➜
New administration set is
attached, reservoir volume
and infusion profi le are reset