30
Section 2
Pump Programming - Intermittent Delivery Mode
Intermittent Programming
Before programming:
>
Stop the pump and change the Lock Level to LL0 as described on
page 9.
>
Select the Intermittent (INTERMT) delivery mode as described in
Changing Delivery Modes on page 10. If the pump is already in the
INTERMT delivery mode, you may select it again to clear all
programming screens to their default settings.
1. Begin at the main screen
>
Make sure the pump is in LL0.
>
Make sure INT and STOPPED
appear on the main screen.
>
Press
„
to begin.
2. Enter the reservoir volume
>
Press
´
or
Î
to select the volume
of fl uid in a fi lled reservoir. (If you
do not wish to use the Reservoir
Volume feature, scroll down to
“Not In Use” located before 1.)
>
Press
€
.
3. Enter the dose volume
>
Press
´
or
Î
to select the
desired dose volume in milliliters.
>
Press
€
.
4. Enter the dose duration
>
Press
´
or
Î
to select the
time period over which dose
will be delivered.
>
Press
€
.
Reservoir Volume
| 100.0 ml
<Range: 1 - 9999>
Dose Volume
| 50.0 ml
<Range: 0.1 - 1000>
*** INT 6240X ***
STOPPED
Press NEXT to advance
Dose Duration
| 1 hrs 00 min
<Range: 00:09-24:00>
IN19822.indd 30
10/24/2011 3:57:16 PM