This guide pertains only to the CADD-Prizm
®
VIP (Variable Infusion
Profile) Model 6100 and 6101 ambulatory infusion pump and the following
delivery modes:
PCA
Continuous
TPN
Intermittent
PCA 6210
CONTIN 6220
TPN 6230
INTERMT 6240
(all revision letters) (all revision letters)
(all revision letters)
(all revision letters)
WARNING:
>
If the pump is dropped or hit, inspect the pump for damage. Do not use a pump
that is damaged or is not functioning properly. Contact Customer Service to
return a pump for service.
>
If the pump is dropped or hit, the battery door may become broken or damaged.
Do not use the pump if the battery door is damaged because the battery will not
be properly secured; this may result in loss of power or non-delivery of drug.
>
Do not disclose to the patient the pump’s security codes or any other information
that would allow the patient complete access to all programming and operating
functions.
>
You must use a CADD
®
extension set with anti-siphon valve or a CADD
®
administration set with either an integral or an add on anti-siphon valve to
protect against unregulated gravity infusion that can result from
an improperly attached cassette.
IN19822.indd 1