Pump Programming - Intermittent Delivery Mode
Section 2
Pump
Pr
ogr
amming
Dose
Volume
KVO
Cycle
Duration
Delay
(using Next
Dose Start
Time)
Time
KVO
Duration
Dose
Dose
Intermittent Cycle Graph and
Programming Screens
The Intermittent delivery mode delivers a specifi ed dose volume
over a specifi ed duration. You may repeat the dose in a cycle of
up to 96 hours. The KVO feature allows you to deliver a minimal
amount of drug between doses to maintain catheter patency. You
may also delay the start of delivery using The next dose start
time feature.
The following fi gure shows the programming screens in the
Intermittent delivery mode:
Air Detector
(Review)
Reservoir Volume
Dose Volume
Dose Duration
Dose Cycle
KVO
Next Dose Start Time
Dose Rate
Milliliters Given
*** INTERMT 6240X ***
STOPPED
Press NEXT to advance
If Air Detector
is installed
IN19822.indd 29
10/24/2011 3:57:15 PM