6
Section 1
General Description and Basic Operations
The Main Screen
The main screen is the starting point for programming or viewing the
pump’s settings. If no keys are pressed, the display will eventually revert
to the main screen. The main screen within each delivery mode displays
information about the delivery status, as shown below.
PCA Delivery Mode
Continuous Delivery Mode
TPN Delivery Mode
Intermittent Delivery Mode
* The Power Source Status will only display when the 9 volt battery is low, unless
the pump has been customized to always show the type of power source in use.
*** PCA 6210X ***
Battery Low RUNNING
Res Vol 50.0 ml
Press NEXT to advance
Power source status*
Delivery mode
RUNNING if the pump is running
DOSING if a demand dose is in
progress
STOPPED if the pump is stopped
RUNNING if the pump is running
STOPPED if the pump is stopped
The current reservoir volume
Reminder that the NEXT key advances to
programming screens
RUNNING
€
if delivery is tapering up
RUNNING
≥
if delivering at the
plateau rate
RUNNING
™
if delivery is tapering down
RUNNING
K
if delivering the KVO
RUNNING
I
if immediately
tapering down
STOPPED if the pump is stopped
DOSING if a dose is in progress
KVO if the KVO rate is in progress
KVO=0 if the pump is running but no
KVO is programmed
STOPPED if the pump is stopped
DELAY if the dose is delayed by next
dose start time (KVO is delivering)
*** CONT 6220X ***
Battery Low RUNNING
Res Vol 50.0 ml
Press NEXT to advance
*** INT 6240X ***
Battery Low DOSING
Res Vol 50.0 ml
Press NEXT to advance
*** TPN 6230X ***
Battery Low RUNNING €
Res Vol 50.0 ml
Press NEXT to advance
IN19822.indd 6
10/24/2011 3:57:09 PM