Resetting the Reservoir Volume
Normally, when you attach a cassette onto the pump as described on
page 36, a series of messages lead you through resetting the reservoir
volume, priming the tubing, (except in LL2), and starting the pump.
You can, however, reset the reservoir volume without changing the
cassette using the reservoir volume programming screen. The pump may
be in any lock level.
NOTE: In the TPN delivery mode, resetting the reservoir volume also resets the
infusion profile so delivery will start at the beginning of the infusion period when
you start the pump.
Resetting reservoir volume without changing the cassette
1. Stop the pump.
2. Press
„
to display the reservoir
volume screen.
3. Press
€
.
4. If this message appears, press
´
to reset the reservoir volume.
(If this message does not appear,
the reservoir volume may already
be reset.)
If this message appears, the pump
is in the TPN delivery mode and
reservoir volume is not in use. Press
´
to reset the infusion profile.
5. If you have reset the reservoir volume
and the pump is in the TPN delivery
mode, you will also see “Infusion
profile is being reset.”
Reservoir Volume
29.2 ml
<Range: Limited>
Reset Reservoir
Volume to 100.0 ml?
Press Y or N
Reset Infusion
Profile?
Press Y or N