46
Section 4
Options
Time
The Time option shows the time of day in 24-hour (military) time according
to the pump’s internal clock. The clock is powered by a separate, internal
battery which retains the time even when the 9-volt battery is removed. The
time is used to record the time of events in the event log.
Date
The Date option should reflect the current date. This feature is used to
record the date of events in the event log.
Air Detector On/Off
The air detector option controls whether the air detector is turned on or
off. This option appears in the menu only if an air detector is installed on
the pump and is not required. (A setting in the biomed toolbox controls
whether an air detector is required. If the air detector is required, you are
not allowed to turn it off and this option will not appear in the menu.)
For certain therapies, it may be desirable to turn off the air detector (for
example, for epidural infusion or subcutaneous infusion).
WARNING:
When the air detector is not installed, or is installed but turned off, the
pump will not detect air in the fluid path. It is recommended that you periodically
inspect the fluid path and remove any air to prevent air embolism.
Event Log, Viewing
The event log records the following types of events: dose delivery, alarms,
errors, power source changes, cassette changes, changes to pump pro-
gramming or settings. The pump records the date and time of each event,
and lists events in order starting from the most recent through the last
500 events or the last delivery mode change.
The pump may be running or stopped and in any lock level to view the
event log.