39
Section 3
Medication Reservoir or Administration Set
Section 3
Reserv
oir or
Admin Set
Priming the Tubing and Connecting to the Patient
If the lock level is LL0 or LL1 when you
attach a cassette, “Prime Tubing?” will
appear in the sequence of screens. Prime
the tubing before connecting it to the
patient’s infusion set or indwelling catheter.
NOTE: If you are not changing the cassette but wish
to prime the fluid path, you may use the Prime
option. To prime, select Prime from the Options menu,
then follow the steps below, beginning with step 2.
WARNING:
Do not prime the fluid path with the
tubing connected to a patient as this could result in
over-delivery of medication or air embolism.
To prime the tubing and connect to the patient
1. When “Prime Tubing?” appears,
press
´
.
2. Make sure the tubing is disconnected
from the patient and the tubing clamp
is open.
3. Press and hold the
´
key until the
tubing is fully primed or until priming
stops.
NOTE: Fluid delivered during priming is
subtracted from the reservoir volume, but is
not added to the Given screen since this fluid
is not delivered to the patient.
4. If the tubing is not yet fully primed,
press
´
and repeat step 3.
When the tubing is fully primed,
press
Î
to exit priming.
Prime Tubing?
Press Y or N
Prime Tubing?
Press Y or N
Disconnect tubing
from patient
Open clamps
Hold Y to prime
Priming...
0.1 ml
Hold Y to prime
Continue Priming?
Press Y or N