Section 2
Pump Programming - PCA Delivery Mode
10. Clear the units given
>
Press
€
if you wish to clear the
total amount delivered since the
time and date shown; even if the
amount is zero, this updates the
time and date markers.
11. Verify the air detector status
This screen will appear only if an air detector is installed.
>
Make sure the setting is correct.
NOTE: If the air detector is not required,
this screen will show whether it is turned
on or off.
>
Press
„
to continue. If you need
to correct the air detector setting,
see the Operator’s Manual.
12. Enter a new patient marker (optional)
This screen will appear only if the extended history is on.
If you do not wish to add a new
patient marker, press
„
.
If you wish to add a new
patient marker to the event log,
>
Press
€
.
>
Press
´
. This will clear the
extended history from the last
patient and add a marker to the
event log. The main screen will
reappear.
13. Review the program
Press
„
repeatedly to review
the programming screens. If you
need to reprogram a setting, press
„
until the appropriate screen
appears and change the setting as
described in this section.
Air Detector
Required
<Review Only>
Clear Extended
History and insert
New Patient Marker?
Press Y or N
To insert New Patient
Marker and clear
Extended History
press ENTER
Milligrams Given
0.00 mg
since 06/08/96 10:35
Press ENTER to clear
IN19822.indd 16
10/24/2011 3:57:11 PM