5
General Description and Basic Operations
Section 1
Basic Oper
ations
Door w/Button»
«Prizm Match
Battery»
Installing a Battery
Use a fresh, 9 volt alkaline or lithium battery such as the DURACELL
®
Alkaline MN 1604 or the ULTRALIFE
®
Lithium U9VL battery. The pump
retains all programmed values while the battery is removed. If the pump
is running, you may connect an external power source to keep the pump
running for 3 minutes while you change the battery.
CAUTION:
>
Do not store the pump for prolonged periods with the battery installed. Battery
leakage could damage the pump.
WARNING:
>
If the pump is dropped or hit, inspect the pump for damage. Do not use a pump
that is damaged or is not functioning properly. Contact Customer Service to
return a pump for service.
>
If the pump is dropped or hit, the battery door may become broken or damaged.
Do not use the pump if the battery door is damaged because the battery will not
be properly secured; this may result in loss of power or non-delivery of drug.
>
There is no pump alarm to alert users that a battery has not been properly
installed or has become dislodged. An improperly installed or dislodged battery
could result in loss of power and non-delivery of drug.
>
Do not use rechargeable NiCad or nickel metal hydride (NiMH) batteries. Do not
use carbon zinc (“heavy duty”) batteries. They do not provide sufficient power
for the pump to operate properly.
To install a battery
1. Make sure the pump is stopped. Press the
button on the battery door and slide the
battery door forward. Remove the
used battery.
2. Match the + and - markings on the new
battery with the markings on the pump.
Insert the battery. The pump will beep if the
battery is inserted correctly.
3. Replace the battery door.
IN19822.indd 5
10/24/2011 3:57:08 PM