23
Pump Programming - Continuous Delivery Mode
Section 2
Pump
Pr
ogr
amming
CONTIN: Programming with Upper Limits,
Adjusting Rate in Lock Level 1
If a prescription allows for the continuous rate to be adjusted during the
course of therapy, you may wish to operate the pump in LL1. Then, when
necessary, you can adjust the continuous rate up to the maximum value
that was programmed in LL0.
Programming the pump to use this feature
1. During initial programming in LL0,
enter the upper limit value for the
continuous rate. (This will be the
maximum value when the pump is
in LL1.)
2. After you are fi nished programming,
change the lock level to LL1.
3. Decrease the continuous rate to
its starting value. “Range: Limited”
indicates you cannot increase the
value beyond the maximum
programmed in LL0.
Adjusting the rate while the pump is in use
If it becomes necessary to increase the continuous rate during the
course of therapy, stop the pump but remain in LL1.
1. Press
„
until the Continuous Rate
screen appears.
2. Press
´
or
Î
to select the desired
value, then press
€
.
“Range: Limited” indicates you
cannot increase the value beyond
the maximum.
3. Restart the pump if appropriate.
Continuous Rate
| 5.0 ml/hr
<Range: 0 - 350.0>
Continuous Rate
| 2.5 ml/hr
<Range: Limited>
Continuous Rate
| 5.0 ml/hr
<Range: Limited>
IN19822.indd 23
10/24/2011 3:57:13 PM