45
Options
Section 4
Options
4. When finished, press
„
to return to the Extended History screen.
NOTE: An asterisk (*) next to a value indicates that it applies only to the selected
time period; it may not match the corresponding value in the programming
screen. For example, “*Given” only reflects doses given during the selected time
period, and may not match the “Given” value on the Dose Counters screen.
Doses Hour by Hour
5. After you select Doses Hour by Hour,
the number of doses given and
attempted during the current hour
will appear.
>
Press
Î
to page back through
through hours.
>
Press
´
to page forward.
6. When finished, press
„
to return to the Extended
History screen.
Changing Delivery Modes
The CADD-Prizm
®
VIP pump contains four delivery modes: PCA,
CONTIN, TPN, INTERMT. To change the delivery mode, see page 10.
When you change the mode, the program will revert to the default
settings, and the event log will clear. Therefore, you may wish to print
the event log before changing the mode. Other Options settings will
not be affected.
AutoLock
The AutoLock Option is available in all delivery modes. It automatically
changes the lock level from LL0 to LL1 or LL2 when the pump is started,
instead of requiring you to manually change the lock level before giving
the pump to the patient. AutoLock may be set to LL1, LL2, or Not In Use.
Doses Hour by Hour
|10:00-10:59 06/08/96
*Given 1
*Attempted 2
IN19822.indd 45
10/24/2011 3:57:19 PM