Pump Programming - PCA Delivery Mode
Section 2
Pump
Pr
ogr
amming
PCA Programming
Before programming:
>
Stop the pump and change the Lock Level to LL0 as described
on page 9.
>
Select the PCA delivery mode as described in Changing Delivery
Modes on page 10. If the pump is already in the PCA delivery mode,
you may select it again to clear all programming screens to their
default settings.
1. Begin at the main screen
>
Make sure the pump is in LL0.
>
Make sure PCA and STOPPED appear
on the main screen.
>
Press
„
to begin.
2. Enter the reservoir volume
>
Press
´
or
Î
to select the volume
of fl uid in a fi lled reservoir. (If you do
not wish to use the reservoir volume
feature, scroll down to “Not In Use”
located before 1.)
>
Press
€
.
3. Enter the units
To accept the current programming units,
press
„
.
Or, to change the units:
>
Press
´
or
Î
to select the desired
programming units.
>
Press
€
.
>
Press
´
to confi rm the change.
NOTE: If the prescription calls for milliliters,
enter milliliters and skip to step 5.
Units
| Milligrams
<Range: mg or ml>
Reservoir Volume
| 100.0 ml
<Range: 1 - 9999>
*** PCA 6210X ***
STOPPED
Press NEXT to advance
Change Units to
Milligrams?
Press Y or N