33
Pump Programming - Intermittent Delivery Mode
Section 2
Pump
Pr
ogr
amming
INTERMT: Stopping the Pump During the Cycle
Stopping the pump during the KVO will not affect the start time of
subsequent doses.
Stopping the pump while a dose is in progress will shift all subsequent
doses by the amount of time the pump is stopped (illustrated below).
If you wish to make up for the lost time in the cycle,
it is recommended
that you wait until the current dose is completed.
To determine the amount
of time remaining in the current dose, use the Time Remaining Option
(described on page 34). Then stop the pump, change to Lock Level 0, and
reprogram the Next Dose Start Time (described on page 34).
KVO
KVO
All subsequent doses
shift by 10 minutes
Pump is stopped
for 10 minutes
Dose
Dose