Section 2
Pump Programming - PCA Delivery Mode
PCA: Starting a Clinician Bolus
1. Make sure the pump is running (in any
lock level). Start the pump if necessary.
2. Press
Œ
.
3. Press
Î
until the clinician bolus
code appears on the display
(See Operator’s Manual).
4. Press
Œ
again.
WARNING:
To prevent the patient from
accessing the clinician bolus function,
do not let the patient know the
clinician bolus code.
5. Press
´
or
Î
to select the
desired amount.
6. Press
€
or
Í
.
NOTE: If you enter a value of 100, a screen
will appear asking you to confi rm the value.
Press
´
to confi rm, or
Î
to re-enter
the value.
7. The screen will show the amount
decreasing as the bolus is delivered.
PCA: Starting a demand dose
If a demand dose has been programmed, the patient may start a
demand dose while the pump is running. If no demand dose has been
programmed, the pump will display the message “Dose not delivered,
No Dose programmed.”
1. Make sure the pump is running
(in any lock level). Start the pump if
necessary.
2. Press
Í
(or the button on the
remote dose cord, if attached). Two
beeps will sound and the pump will
begin delivering the demand dose.
As the demand dose is delivered, the
main screen will show “DOSING” in
place of “RUNNING.”
Clinician Bolus Code
XXX
Clinician Bolus
| 10.00 mg
Range <0 - 20.00>
Demand Dose
Started
NEXT to continue
*** PCA 6210X ***
Low Battery DOSING
Res Vol 47.0 ml
Press NEXT to advance
Clinician Bolus
10.00 mg
<Delivering...>
IN19822.indd 18
10/24/2011 3:57:11 PM