Instrument overview
Instruction Manual ScrubberGuard
14830E/2
19
2.4
Technical data for the ScrubberGuard
2.4.1
General technical data
ScrubberGuard
Values
Sample temperature
▪
0 .. +50 °C (not > 30 °C above ambient temperature)
▪
0 .. +60 °C (with optional solenoid valve)
Sample inlet pressure
▪
0 .. 3 bar
▪
0 .. 4.5 bar (with optional pressure reducing valve)
▪
-0.7 .. 0 bar (with optional sample inlet pump)
The pressure in the return line must not exceed 3 bar.
Sample flow
4 to 20 l/min
Ambient temperature
Max. 50 °C
Humidity
0 .. 100 % relative humidity, non-condensing
Protection class
IP54
Power supply
215 .. 240 VAC, 50 / 60 Hz
Power consumption
650 W / 1000 VA (1000 W / 1700 VA incl. inlet pump)
Interfaces
Ethernet (RJ45 connector)
Inclination/heel
Operation up to 30° and reliable measurement up to 20° incli-
nation in all axes.
Weight
Approx. 100 kg
Materials
▪
Frame / structure: 316L
▪
In contact with medium: 1.4462, 316L, PVC-U, FKM, NBR,
polycarbonate
▪
Casing: Aluminum, powder-coated
Free-fall measuring cell
material
▪
Inlet pipe: stainless steel 1.4435
▪
Outlet: PVC
Display on AquaScat
SG / OilGuard SG
¼ VGA with touchscreen
Resolution: 320 x 240 pixels with 3.5" diagonal
Operation
Touchscreen
Outputs/inputs
Outputs:
▪
4 x 0/4 .. 20 mA, galvanically isolated up to max. 50 V rela-
tive to ground, max. 500 Ω burden
▪
2 x relay contacts 30 V, 2 A
Inputs:
▪
1 x digital input (for external NO contact)
Dimensions
Approx. 1280 x 880 x 400 mm (W x H x D)
Connections: Electrical
cable cross-sections
0.25 .. 4 mm
2
, AWG 22 .. 12
Hydraulic connection
R1"