Servicing
Instruction Manual ScrubberGuard
14830E/2
101
Remove the optics unit as follows and place it in the recess (Dockingstation):
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Open the mounting clips (circles in figure be-
low) as follows:
Use a little force to push the red safety catch
in the direction of the arrow (figure 1) and at
the same time lift the mounting clip (figure
2). Press the mounting clip in the direction of
the arrow over the lock plate of the optics
unit (figure 3) and then open (figure 4).
Position of the mounting clips
Figure 1
Figure 2
Figure 3
Figure 4
2.
Remove the optics unit (Figure 32, pos. 2)
from the measuring cell unit (Figure 32, pos.
1) and position it in the recess (Dockingsta-
tion) (Figure 32, pos. 3).
The tolerance of the cutout prevents
the instrument from tipping over.
9.4.2
Checking the flow rate and cleanliness of the measuring instru-
ments
The following procedure describes how the flow rate is checked on the photometers:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the optics unit on the photometer
according to Section 9.4.1 and place in the
recess (docking station).
Danger due to UV radia-
tion (on OilGuard SG only).
Use UV goggles and gloves.