Instruction Manual ScrubberGuard
Troubleshooting
110
14830E/2
10
Troubleshooting
10.1
Pinpointing faults
DETECTABLE FAULT
MEASURE
No reading
▪
Check whether the supply voltage is connected.
▪
Check whether the fine-wire fuse is OK (Reference
Handbook).
Fault message in the display
▪
Analyze the fault message according to Section
10.2 to Section 10.4.
The reading appears to be
wrong
▪
Ensure that the sample to be measured corresponds
to the operating conditions. Section 2.4
▪
Ensure that the sample is free from air bubbles.
Section 9.4.2
▪
Carry out recalibration. Section 9.4.3
▪
Check whether the system is correctly mounted.
Section 4
▪
Ensure that the servicing duties have been perfor-
med according to the servicing schedule. Section 9
Table 2: Pinpointing malfunctions
If the listed measures do not lead to the desired results, please consult customer service.
Section 12
10.2
Warning messages and effect on operation
Warnings indicate an unusual state.
WARNINGS
If a warning occurs during operation, it has the follo-
wing effects:
▪
The system continues to operate. However, the
measuring results must be evaluated with cau-
tion. The cause of the warning message should
be remedied at the next possible opportunity.
▪
When the cause of the warning has been
remedied, it is automatically deleted.
▪
When the
Warning
message
occurs, the color of
the status display changes to
orange
and the
warning text describes the warning in question.
Example:
WARNING HUMIDITY