Servicing
Instruction Manual ScrubberGuard
14830E/2
99
WORKSTEP
ADDITIONAL INFO / IMAGES
4.
Select the
Hamilton
menu.
Press
start…
under the “Find Sensor” menu
item.
5.
A system search is now carried out for con-
nected Hamilton sensors. As soon as a sensor
is found, the type and slave number are dis-
played (e.g. Oxygen, Slave no. 1).
If the slave number has not yet been adjusted
to Siginet, then
undefined
is shown under
the slave number menu item. The slave num-
ber must be set to 2.
6.
Select the
Siginet
menu and press
start…
under “Network scan”.
A search is made for a few seconds and then
a list appears with all found sensors.
If not all sensors are displayed, then proceed
as follows:
1. Start the
“Network scan”
again.
2. Check the connections to the sensors.
3. Check whether each sensor has an indivi-
dual slave number.
7.
The sequence must be as follows:
1.
S1 OilGuard
2.
S2 pH
If the sequence is not as desired, then the
slave numbers can be reassigned. To do this,
select the sensors in the desired sequence.
The new slave number is shown and the asso-
ciated button changes to green.
If the sequence of the sensors is correct, con-
firm with the
OK
button.
8.
The system can be put into operation.