Servicing
Instruction Manual ScrubberGuard
14830E/2
97
WORKSTEP
ADDITIONAL INFO / IMAGES
5.
Carrying out recalibration for nominal
value 1:
Compare the
Nominal value
(circle) with the
value on the calibration solution.
Pressing the
Nom. val.
button (circle)
opens a numeric input field where the nomi-
nal value can be adjusted.
6.
Wait until the temperature (circle) is stable.
Recalibration is only carried out when
the values have been stable for the past
three minutes.
7.
Press the
initiate...
button. Recalibration
starts.
If the adjustment was successful, this is con-
firmed with
Adjustment OK
. This completes
the adjustment.
If the adjustment was not successful, the fol-
lowing messages may appear:
running…
Cause: Values not yet stable.
Diff. to small
Cause: The nominal values of the calibration
solutions are too close together.
Measure:
▪
Correct calibration solution used?
▪
Calibration solution OK?
If the displayed quality is
between 100 and 35 following
calibration, then this is due to
aging of the sensor.
If the calibration was incorrect,
then a quality value of 30 is dis-
played. Calibration must be re-
peated for both nominal
values.
Out of tolerance
Cause: Current actual value is too far away
from the nominal value.
Measure:
▪
Check that the set nominal value matches
the nominal value of the calibration solu-
tion.
▪
Clean the sensor.