Servicing
Instruction Manual ScrubberGuard
14830E/2
69
WORKSTEP
ADDITIONAL INFO / IMAGES
2.
Visual inspection of the deaeration tubes:
▪
Is there an excessive deposit build-up on
the glass of the deaeration tube?
▪
Is there an absence of water/air flowing
through the respective deaeration hose?
▪
Is the spout in the deaeration tube blo-
cked (arrow)?
▪
Are there leaky connections?
▪
Are the deaeration tubes damaged?
If any of these questions are
answered in the affirmative, please
consult the Section 9.2.4.
3.
Empty and clean the siphon on the pH sensor.
To do this, remove the sealing cap (arrow).
To prevent the pH sensor from drying
out, it is essential that the siphon is filled
with water and locked.
4.
Visual inspection of the water tank:
▪
Is the inside of the water tank dirty?
▪
Are there indications of leaky connec-
tions?
▪
Is the float switch inserted incorrectly (see
figure)?
▪
Is there rust on the water tank?
If any of these questions are
answered in the affirmative, please
consult the Section 9.2.3.