Servicing
Instruction Manual ScrubberGuard
14830E/2
107
WORKSTEP
ADDITIONAL INFO / IMAGES
7.
Put the outlet cone C onto the outlet pipe E
up to the stop. When doing so, hold the
outlet pipe from below. Now press the out-
let cone together with the outlet pipe down-
ward up to the stop.
8.
Fasten this unit by tightening the conduit
gland.
At position X, there should now be no space
between the outlet cone and the conduit
gland!
9.
Insert the inlet pipe into the measuring cell
unit up to the stop and then fasten in place
with the screw connections X and Y.
10. Place the optics unit back on the measuring
cell unit and lock with the four mounting
clips.
Pay attention to the guide pins (see figure).