Servicing
Instruction Manual ScrubberGuard
14830E/2
67
9
Servicing
CAUTION!
Damage to the instrument due to servicing duties that have not been carried out
or have been carried out improperly.
If servicing duties are not carried out according to the servicing schedule or non-original
SIGRIST spare parts are used, this can lead to damage to the instrument or measuring er-
rors.
In this case, SIGRIST-PHOTOMETER AG accepts no warranty claims made by the customer
and is not responsible for any subsequent costs. To avoid this situation, please adhere to
the following steps:
▪
Carry out servicing duties according to the servicing schedule (Section 9.1).
▪
When carrying out servicing duties, use original SIGRIST spare parts according to the
spare parts list (Section 16). The use of third-party spare parts requires the written ap-
proval of SIGRIST-PHOTOMETER AG.
▪
If the instruments are subjected to heavy use or exposed to severe environmental con-
ditions, servicing duties must be carried out more often as required. Wear parts in par-
ticular must be replaced more often depending on the circumstances.
9.1
Servicing schedule for the ScrubberGuard
Servicing duties are obligatory measures for maintaining functional efficiency.
WHEN
WHAT
WHERE
WHO
Monthly
Inspection using checklist according to
Section 9.2.1
System
User
Check the flow rate on the system ac-
cording to Section 9.2.2
System
User
Every 3 mon-
ths
Replace the impeller on the inlet / ou-
tlet pump according to Section 9.2.8
System
User
Clean and calibrate the pH sensor ac-
cording to Section 9.3.2
pH sensor
User
Perform manual adjustment according
to Section 9.4.3
AS SG
OG SG
Annually
Replace the air filter according to Sec-
tion 9.4.5
AS SG (A)
OG SG (A)
User
Replace the float switch according to
Section 9.2.7
System
Replace the impeller set / seal set on
the inlet / outlet pump according to
Section 9.2.8