Operation
Instruction Manual ScrubberGuard
14830E/2
55
7.11
Switching to service mode
The system is configured in service mode. The measuring procedure is interrupted and the
main menus appear on the display. Proceed as follows to access service mode:
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Enter the access code and confirm with
OK
.
Factory setting is
0
.
3.
Select menu
Local
…….
or
S 1 .. 8
.
Now the instrument is in service
mode.
The following applies in service mode:
▪
* The measuring values remain on the last values on the digital interfaces.
▪
* Depending on the configuration, the current outputs go to 0/4 mA or remain on the
last measuring value.
▪
The limits are deactivated.
▪
If an output for service is programmed, it is activated.
▪
Error messages are suppressed.
* This applies when the
Local parameters\Current outputs\General\For service
is set to
Measure
.
For measuring mode press the
Meas
button. When switching from service mode to meas-
uring mode, an hourglass appears in the information bar for about 20 seconds. The meas-
uring values are frozen during this time.