Settings
Instruction Manual ScrubberGuard
14830E/2
61
8.3
Setting the limits
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Enter the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Local ……
button.
If the desired menu does not
appear, press the arrow at the bot-
tom right.
4.
Press the
Limits
button.
5.
Select
Limit 1 .. 8
.
6.
Select the source of the measuring channel
from the
Source
menu item. This name is dis-
played to simplify identification of the meas-
uring channel.
The channels defined under
Meas.channels
, three math chan-
nels and two analog channels are
available for selection.
→
Reference Handbook
7.
Define the
Mode
.
The following selection is available:
▪
Inactive
(limit monitoring of
this channel is deactivated)
▪
Exceeded
(limit active when
the set threshold is exceeded)
▪
Undershot.
(limit active when
the set threshold is undershot)
8.
Define the upper limit, lower limit, cut-in de-
lay and cut-out delay with the number pad.
Pressing the current number
value takes you to the entry mode.
9.
Press the
Meas
button.
The instrument is in measuring
mode again.
The limits have to be configured accordingly so that they are not only displayed, but that the
outputs are also switched.