Instruction Manual ScrubberGuard
Settings
60
14830E/2
8.2
Setting the current outputs
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Enter the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Local ……
button.
4.
Press the
Curr. outputs
button.
If the desired menu does not
appear, press the arrow at the bot-
tom right.
5.
Select
Current 1 .. 4
(
1 .. 8
).
6.
Select the source of the measuring channel
from the
Source
menu item. This name is dis-
played to simplify identification of the measu-
ring channel.
The channels defined under
Meas.channels
, three math chan-
nels and two analog channels are
available for selection.
→
Refe-
rence Handbook
7.
Select the
Range
.
MR1 .. MR8 (see table below)
In 1, In 2, Auto 1, Auto 2 (
→
Refe-
rence Handbook)
8.
Press the
Meas
button.
The instrument is in measuring
mode again.
Measuring range no.
Measuring range
(standard)
Measuring range
(customer-specific)
MR1
0 .. 1000
MR2
0 .. 500
MR3
0 .. 200
MR4
0 .. 100
MR5
0 .. 50
MR6
0 .. 25
MR7
0 .. 14
MR8
0 .. 10
If other measuring ranges are needed, you can re-program the table above as required.
→
Reference Handbook