Instruction Manual ColorPlus 2
Servicing
92
13045E/3
WORKSTEP
ADDITIONAL INFO / IMAGES
5.
The current measuring value on channel 1 is
displayed in the
Recalibration\C1
menu under
Act. val.
Depending on the setting in
Meas. chan-
nels\Channel n\Lin/Log
the measuring value
is shown in % transmittance (
Lin
) or in ab-
sorbance (
Log
).
Compare the actual value with the value on
the checking unit.
In the event of a deviation of more than ±2%
the measuring value or absolute 0.1%T, repeat
steps 1 to 5.
If the measured value deviates once again,
then contact the service center.
On instruments with several channels, press
the
ESC
button and then compare the actual
value and nominal value on all other channels.
If the measured value matches the value on
the housing, then continue from point 6.
6.
Remove the checking unit and reattach the
cover.
The instrument can now be operated again.