Servicing
Instruction Manual ColorPlus 2
13045E/3
89
9.10
Recalibrating the photometer with sliding measuring cell
Recalibrating the photometer can result in deviations from the previous measuring value as
the instrument is newly reset to a reference value (e.g. distilled water).
In order for recalibration with the sliding measuring cell to be carried out, the instrument
must be specially configured at the factory. After this configuration has been made,
Calibr.
cell
is displayed in the title bar of the
Recalibr./Channel
menu.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Loosen the lock (X) and slide the measuring
cell from the measurement position.
2.
Insert the holder for the funnel in the bore (ar-
row).
3.
Insert the calibration measuring cell up to the
stop in the guides, then hang the funnel in the
holder.