Servicing
Instruction Manual ColorPlus 2
13045E/3
85
WORKSTEP
ADDITIONAL INFO / IMAGES
3.
Clean the measuring cell window with a paper
towel, cloth or warm, soapy water.
Replace the seal or measuring cell win-
dow, if required.
4.
Position the seal (Figure 30, pos. 2) in the bore
of the calibration measuring cell.
5.
Reposition the measuring cell window (Figure
30, pos. 3) on the seal.
6.
Position the Teflon slide ring (Figure 30, pos. 4)
on the measuring cell window.
7
Screw in the ring nut (Figure 30, pos. 5) with
the special wrench.
Do not fasten too tightly, as the measur-
ing cell windows can break.