Servicing
Instruction Manual ColorPlus 2
13045E/3
83
WORKSTEP
ADDITIONAL INFO / IMAGES
3.
Slide the photometer from the measurement
position.
4.
Remove the ring nut (Figure 29, pos. 5) using
the face wrench.
5.
Remove the Teflon slide ring (Figure 29, pos.
4), measuring cell window (Figure 29, pos. 3)
and the seal (Figure 29, pos. 2) from the bore.
6.
Clean the measuring cell window with a paper
towel, cloth or warm, soapy water.
Replace the seal or measuring cell win-
dow, if required.
7.
Position the seal (Figure 29, pos. 2) in the bore.
8.
Reposition the measuring cell window (Figure
29, pos. 3) in the bore.
9.
Position the Teflon slide ring (Figure 29, pos. 4)
on the measuring cell window.
10. Screw in the ring nut (Figure 29, pos. 5) with
the special wrench.
Do not fasten too tightly, as the measur-
ing cell windows can break.
11.
Slide the instrument back into the measure-
ment position and fasten in place with the
locking screw.
12. Carry out recalibration according to
Section 9.9.