Instruction Manual ColorPlus 2
Servicing
94
13045E/3
WORKSTEP
ADDITIONAL INFO / IMAGES
6.
Pull the UV light source from the holder.
7.
Insert the new UV light source up to the stop
in the holder and fasten in place with the Allen
bolt.
The marking on the light source must
match that on the holder.
8.
Reinsert the plug on the print.
9.
Replace the desiccant bag according to Section
9.7.
10
Reassemble the instrument in reverse order.
11. Carry out recalibration according to
Section 9.9.