Instruction Manual ColorPlus 2
Servicing
78
13045E/3
WORKSTEP
ADDITIONAL INFO / IMAGES
5.
Clean the measuring cell windows on the
transmitter/receiver side.
If cleaning of the measuring cell windows was
successful
, then continue from
point 11
.
If cleaning of the measuring cell windows was
not successful
, then continue from
point 6
.
6.
Remove the screw connection on the measur-
ing cell with the special wrench (32 mm).
7.
The screw connection, pressure ring, seal and
measuring cell windows can now be removed.
8.
Clean the measuring cell windows with a suit-
able cleaning agent.
Ensure that the measuring cell windows are no
longer soiled during the subsequent assembly.
Replace the measuring cell window and seal, if
required.
9.
Replace the seal on the flood protection (A), if
required.
10. Reassemble the instrument in reverse order.
11. Carry out recalibration according to
Section 9.9.