Instruction Manual ColorPlus 2
Servicing
86
13045E/3
9.7
Replacing the desiccant on the transmitter
DANGER!
Danger of injuries on the UV light source caused by electric shocks, UV radiation
and high temperatures.
When replacing the desiccant, careless handling can lead to electric shocks, eye injuries
caused by UV radiation and burns due to temperatures in excess of 80 °C. It is thus essen-
tial to observe the following safety measures when removing the UV light source:
The service voltage must be interrupted before opening the housing.
The instrument must not be operated when the housing is removed under any circum-
stances.
Before opening, wait until the instrument has cooled down to a manageable tempera-
ture.
The following procedure describes how to replace the desiccant on the transmitter:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the service voltage to the photo-
meter.
2.
Loosen the three hex bolts and remove the
housing.
3.
Replace the desiccant as follows:
Remove the old, saturated desiccant bag (A).
On the new desiccant bag, shake and roll up
the contents at one end.
The new desiccant can then be used.
4.
Check the housing seal (B) and replace, if re-
quired.
Check also the cover screw seals and replace, if
required.
5.
Reassemble the instrument in reverse order.