Instruction Manual ColorPlus 2
Servicing
84
13045E/3
9.6
Cleaning or replacing the windows on the calibration
measuring cell
Figure 30: Calibration measuring cell window
Calibration measuring cell
Neoprene seal
Measuring cell window
Teflon slide ring
Ring nut
The cleaning or replacement process is the same for both calibration measuring cell win-
dows. The following steps describe the procedure for only one window.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the ring nut (Figure 30, pos. 5) using
the face wrench.
2.
Remove the Teflon slide ring (Figure 30, pos.
4), measuring cell window (Figure 30, pos. 3)
and the seal (Figure 30, pos. 2) from the cali-
bration measuring cell.