Servicing
Instruction Manual ColorPlus 2
13045E/3
81
WORKSTEP
ADDITIONAL INFO / IMAGES
13. Position the seal (C) in the groove of the flood
protection.
If required, replace the seal (C) of the
flood protection.
14. Attach the ring adapter (B) (including attached
OPL bit) on the transmitter (A) and receiver (C)
and fasten in place with the four hex bolts.
15. Clean all contact surfaces (arrows) on the inline
housing.
16. Reinsert the transmitter and receiver with at-
tached ring adapter and OPL bit into the inline
housing.
The groove (X) on the transmitter and
receiver must face forwards when mounted.
17. Fasten the transmitter and receiver back onto
the inline housing using the lock rings.
18. Carry out recalibration according to
Section 9.9.