Mounting
Instruction Manual ColorPlus 2
13045E/3
31
4.2
Mounting the photometer on the inline housing
4.2.1
Mounting position of the photometer in the inline housing
Using a standardized inline housing (VARINLINE
®
or compatible models), the photometer can
be installed in both horizontal and vertical product lines.
Correct installation position in a
vertical
product line
Position X: Position of the groove
Correct installation position in a
horizontal
product line
Incorrect
installation position in a
horizon-
tal
product line
The following points must be observed here:
The groove (X) on the transmitter and receiver must face forwards when mounted.
Always mount the photometer so that the transmitter and receiver are horizontally op-
posite one another. Accordingly, only install in a housing where this layout is possible.
The photometer must be mounted in the line at least 2 meters away from inspection
glasses and other unwanted light sources.