Servicing
Instruction Manual ColorPlus 2
13045E/3
87
9.8
Replacing the desiccant on the receiver
The following procedure describes how to replace the desiccant on the receiver:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the service voltage to the photome-
ter.
2.
Loosen the three hex bolts and remove the
cover on the receiver.
3.
Replace the desiccant as follows:
Remove the old, saturated desiccant bag (A).
On the new desiccant bag, shake and roll up
the contents at one end.
The new desiccant can then be used.
4.
Check the housing seal (B) and replace, if re-
quired.
Check also the cover screw seals and replace, if
required.
5.
Reassemble the instrument in reverse order.