Instruction Manual ColorPlus 2
Servicing
82
13045E/3
9.5
Cleaning or replacing the measuring cell windows on the
sliding measuring cell
Figure 29: Cleaning the measuring cell windows, sliding measuring cell
Sliding measuring cell
Kalrez seal, or other
Measuring cell window
Teflon slide ring
Ring nut
The cleaning or replacement process is the same for both measuring cell windows. The fol-
lowing steps describe the procedure for only one window.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Stop the sample flow and let the measuring
cell run dry.
2.
Loosen the lock (X).