Troubleshooting
10.1
Pinpointing malfunctions
DETECTABLE FAULT
MEASURE
No reading
Check whether the supply voltage is connected.
Check whether the mains plug is connected.
Check whether the instrument is switched on.
Fault message in the display
Analyze the fault message according to Section 10.3.
The reading appears to be
wrong
Ensure that the sample to be measured corresponds
to the operating conditions (Section 2.9.1/
Section 9.9).
Carry out recalibration ( Section 9.10).
Check whether the instrument is correctly mounted
(Section 4).
Ensure that the servicing duties have been performed
according to the servicing schedule (Section 9.1).
Table 2: Pinpointing malfunctions
If the listed measures do not lead to the desired results, please consult customer service (Sec-
tion 11).